Digital Multimeter

DME1700 Series

Price range: 158,000円 through 198,000円

  • Resolution : 6 1/2 digit
  • Enhanced visibility: 4.3-inch TFT color LCD
  • Basic measurement function
    • DC voltage: 100 mV, 1 V, 10 V, 100 V, 1000 V
    • AC voltage: 100 mV, 1 V, 10 V, 100 V, 750 V
    • DC current: 10 mA, 100 mA, 1 A, 3 A
    • AC current: 10 mA, 100 mA, 1 A, 3 A
    • 2-wire/4-wire resistance: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ (100 MΩ range is 2-wire resistance only)
    • Frequency and period measurement: 3 Hz to 300 kHz
    • Other measurement functions: Continuity test, diode test, temperature measurement
    • Measurement range: Selectable in auto range or manual range
    • Maximum sample count: 1,500,000 (DME1701 only)
  • Standard communication interfaces included (DME1701)
    LAN, USB *Options: GPIB, RS232C (selected at time of order)

ImageSKUInterfaceMaximum reading speed標準価格WEBかんたんお見積り
DME1700 SeriesDME1700–300 rdg/s158,000円
(税込:173,800円)
dme1700シリーズDME1701USB, LAN10000 rdg/s198,000円
(税込:217,800円)
SKU: DME1700 Series

Line-up

ImageSKUInterfaceMaximum reading speed標準価格WEBかんたんお見積り
DME1700 SeriesDME1700–300 rdg/s158,000円
(税込:173,800円)
dme1700シリーズDME1701USB, LAN10000 rdg/s198,000円
(税込:217,800円)

Product Overview

Precise measurements and quick readings for various testing requirements.
Measurement data can be exported directly to a USB flash drive as a CSV file.

The DME1700 Series digital multimeters have 6.‑digit resolution. This series provides various measurement functions, including voltage, current, resistance, frequency, and temperature, making it suitable for many evaluation and measurement applications.
This series features a 4.3‑inch TFT color LCD. Auto-ranging provides user‑friendly operation. Moreover, DME1700 achieves an accuracy of 0.0035% for DC voltage (10 V range) and 0.01% for resistance (10 kΩ range)*. It delivers dependable measurements in environments that require high precision.

*Value for the DME1701 within one year of calibration.

At the fastest integration time (PLC), the DME1700 achieves a maximum of 300 rdg/s (0.2 PLC) and the DME1701 a maximum of 10,000 rdg/s (0.006 PLC), which is ideal for automated measurement systems that require fast, high-volume measurements.
DME1701 comes standard with LAN. USB, GPIB* and RS-232C* interfaces are available as options, providing flexibility for embedded applications. When used with an optional scanner card, the unit can measure up to 20 channels, making it well-suited for multi-point evaluation.

*Selected at time of order.

Specifications

DC voltage measurement (DCV)

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionInput resistanceAccuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
100 mV0.1 μV>10 GΩ0.0090+0.00650.0005+0.00050.0050+0.00350.0005+0.0005
1 V1.0 μV0.0080+0.00100.0005+0.00010.0040+0.00070.0005+0.0001
10 V10 μV0.0075+0.00050.0005+0.00010.0035+0.00050.0005+0.0001
100 V100 μV10 MΩ0.0085+0.00060.0010+0.00010.0060+0.00060.0010+0.0001
1000 V1 mV0.0085+0.00100.0010+0.00010.0060+0.00100.0010+0.0001

DC current measurement (DCI)

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionShunt resistanceAccuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
10 mA10 nA5.1 Ω0.050+0.0200.0020+0.00200.050+0.0200.0020+0.0020
100 mA100 nA0.050+0.0050.0020+0.00050.050+0.0050.0020+0.0005
1 A1 μA0.1 Ω0.100+0.0100.0050+0.00100.100+0.0100.0050+0.0010
3 A10 μA0.200+0.0020.0050+0.00200.200+0.0200.0050+0.0020

2-wire/4-wire resistance measurement (Ω2W/Ω4W)

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionTest currentAccuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
100 Ω100 μΩ1 mA0.014+0.0070.0006+0.00050.010+0.0040.0006+0.0005
1 kΩ1 MΩ0.014+0.0010.0006+0.00010.010+0.0010.0006+0.0001
10 kΩ10 mΩ100 μA0.014+0.0010.0006+0.00010.010+0.0010.0006+0.0001
100 kΩ100 mΩ10 μA0.014+0.0010.0006+0.00010.010+0.0010.0006+0.0001
1 MΩ1 Ω0.014+0.0010.0010+0.00020.010+0.0010.0010+0.0002
10 MΩ10 Ω1 μA0.040+0.0010.0030+0.00040.040+0.0010.0030+0.0004
≧ 100 MΩ*1100 Ω1 μA | 10 MΩ0.800+0.0100.1500+0.00020.800+0.0100.1500+0.0002

*1 The ≥ 100 MΩ range is only available for 2-wire resistance measurement (Ω2W).


Diode test

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionTest current (±5 %)Accuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
10V (fixed)10 μV1 mA0.01+0.030.001+0.0020.01+0.030.001+0.002

Continuity test

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionTest current (±5 %)Accuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
1 kΩ (fixed)10 mΩ1 mA0.01+0.030.001+0.0020.01+0.030.001+0.002

DC measurement characteristics

Open in a new window
ItemDME1700DME1701
DC voltage measurement
(DCV)
Over-rangeFor 100 mV to 100 V ranges20 % over-range allowed
For 1000 V range1 % over-range allowed
A/D converter linearityFor 10 V range0.0001 % of measurement range
Input impedance selection *1For 100 mV to 10 V ranges10 MΩ ± 1 % or >10 GΩ selectable
For 100 V to 1000 V range10 MΩ ± 1 % only
Input bias currentAt 23 °C (73.4 °F)Less than 50 pA
Common mode voltage (Between LO terminal and chassis)Max. 500 Vpk
Auto zero adjustmentWithin ± 1 °C (± 1.8 °F) and within 10 min of the last auto zero adjustmentAdding ± (0.0002 % of measurement range + 3 μV) to accuracy
additional error for disabled settingWithin ± 5 °C (± 9 °F) and within 60 min after the last auto zero adjustmentAdding ± (0.001 % of measurement range + 10 μV) to accuracy
DC current measurement
(DCI)
Over-rangeFor 10 mA to 1 A ranges20 % over-range allowed
For 3 A rangeNo over-range
Common mode current50 Hz or 60 Hz< 600 nApk-pk

*1 Specifications for parallel connection of less than 400 pF.


Frequency and period measurement (FREQ/PERIOD)

Open in a new window
Input signal rangeFrequency measurement rangeDME1700DME1701
Accuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
100 mV to 750 V
(10 mA to 3 A)
3 Hz to 10 Hz0.10.0050.10.1
10 Hz to 100 Hz0.050.0050.030.035
100 Hz to 1 kHz0.020.0010.010.015
1 kHz to 300 kHz0.0120.0010.010.015

Measurement characteristics (over-range)

Open in a new window

Measurement characteristics (over-range)

ItemDME1700DME1701
Frequency and Period measurement
(FREQ/PERIOD)
For ACV input (all measurement ranges)Shared with AC voltage measurement (ACV)
For ACI input (all measurement ranges)Shared with AC current measurement (ACI)
Continuity test (CONT)1000 Ω range (fixed)1 % over-range allowed
Diode testAll measurement rangesNo over-range

AC voltage measurement (ACV)

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionFrequencyAccuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
100 mV to 750 V0.1 μV to 1 mV3 Hz to 5 Hz1.00+0.030.100+0.0031.00+0.040.100+0.003
5 Hz to 10 Hz0.38+0.030.035+0.0030.35+0.040.035+0.003
10 Hz to 20 kHz0.09+0.030.005+0.0030.06+0.040.005+0.003
20 kHz to 50 kHz0.15+0.050.011+0.0050.12+0.050.011+0.005
50 kHz to 100 kHz0.63+0.080.060+0.0080.60+0.080.060+0.008
100 kHz to 300 kHz4.00+0.500.200+0.0204.00+0.500.200+0.020

AC current measurement (ACI)

Open in a new window
ItemDME1700DME1701
Measurement rangeResolutionFrequencyAccuracy (1 year)Temperature coefficient/°CAccuracy (1 year)Temperature coefficient/°C
10 mA10 nA3 Hz to 5 Hz1.00+0.040.100+0.0061.00+0.040.100+0.006
5 Hz to 10 Hz0.30+0.040.035+0.0060.30+0.040.035+0.006
10 Hz to 5 kHz0.10+0.040.015+0.0060.10+0.040.015+0.006
5 kHz to 10 kHz (TYP)0.10+0.040.030+0.0060.10+0.040.030+0.006
100 mA100 nA3 Hz to 5 Hz1.00+0.040.100+0.0061.00+0.040.100+0.006
5 Hz to 10 Hz0.30+0.040.035+0.0060.30+0.040.035+0.006
10 Hz to 5 kHz0.10+0.040.015+0.0060.10+0.040.015+0.006
5 kHz to 10 kHz (TYP)0.10+0.040.030+0.0060.10+0.040.030+0.006
1 A1 μA3 Hz to 5 Hz1.00+0.061.000+0.0601.00+0.061.000+0.060
5 Hz to 10 Hz0.35+0.040.035+0.0060.35+0.040.035+0.006
10 Hz to 5 kHz0.23+0.040.015+0.0060.23+0.040.015+0.006
3 A10 μA3 Hz to 5 Hz1.00+0.061.000+0.0601.00+0.061.000+0.060
5 Hz to 10 Hz0.35+0.040.035+0.0060.35+0.040.035+0.006
10 Hz to 5 kHz0.23+0.040.015+0.0060.23+0.040.015+0.006

AC measurement characteristics

Open in a new window
ItemDME1700DME1701
Common specificationsAC measurement modeWith anti-aliasing filter AC-coupled digital sampling
Crest factor
(except sine waves)
Crest factor selectionThe larger of the following crest factors:
• Up to 3:1 of full scale input • Max. 10:1
Measurement range selection in auto rangeOptimal measurement range for the crest factor up to 10:1
Applying the accuracy characteristics[Crest factor] × [fundamental frequency] ≤ 3 kHz
(applied to all crest factors)
Volt-Hertz product8 × 107 V*Hz or below
Common mode rejection ratio (If LO terminal lead imbalance is 1 kΩ)>>70 dB
AC filterMeasuring input signals with frequencies greater than the set AC filter bandwidth
Effect of AC filter bandwidth on aperture time• 2 ms at 200 Hz
• 20 ms at 20 Hz
• 200 ms at 3 Hz
Input impedance (if parallel connection < 100 pF)1.1 MΩ ± 2 %
Measurement range additional errorIf the input is below 50 kHz and is 1 % to 5 % of the measurement rangeAdding 0.1 % of the measurement range to the accuracy
If the input is 50 kHz to 100 kHzAdding 0.13 % of the measurement range to the accuracy
AC voltage measurement
(ACV)
Over-rangeFor 100 mV to 100 V ranges20 % over-range allowed
For 750 V rangeNo over-range
Maximum DC voltageAll ACV measurement ranges400 V
AC current measurement
(ACI)
Over-rangeFor 10 mA to 1 A ranges20 % over-range allowed
For 3 A rangeNo over-range

General Specifications

Open in a new window
ItemDME1700DME1701
Nominal voltage range (allowable voltage range)100 Vac to 240 Vac (90 Vac to 264 Vac)
Power consumption25 VAmax
Environmental conditionsOperating temperature range / Operating humidity range0 °C to 50 °C (32 °F to 122 °F) / 80 %rh or less (no condensation)
Storage temperature range-40 °C to 70 °C (-40 °F to 158 °F)
AltitudeUp to 2000 m
Dimensions / Weight212.8(8.38)W × 88.3(3.48)H× 272.3(10.72)D mm(inches) (Main unit only. Protrusions and other parts are not included) / 2.5 kg (5.5 lbs)
Interface–USB, LAN
Safety*1Complies with the requirements of the following directive and standards.
Low Voltage Directive 2014/35/EU*4
EN 61010-1 (Class I *2, Pollution Degree 2 *3), EN 61010-2-030
Measurement Category II 300 V
EN 61010-031
Electromagnetic compatibility (EMC)*1 *4Complies with the requirements of the following directive and standards.
EMC Directive 2014/30/EU, EN 61326-1, EN 61326-2-1,
EMI: CISPR 11 (Class A *5, Group I *6),
EN 61000-3-2, EN 61000-3-3
EMS: EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5,
EN 61000-4-6, EN 61000-4-8, EN 61000-4-11
Applicable under the following conditions: The maximum length of all cabling and wiring connected to the product must be less than 3 m.
Input protectionUT HI terminal1000 Vpk
SENSE HI terminal350 Vpk (referenced to LO terminal)
SENSE LO terminal
Insulation resistance (Between each terminal and chassis)500 Vpk, >10 GΩ
Capacity (Between each terminal and chassis)< 300 pF

*1 Does not apply to specially ordered or modified products.
*2 This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded.
*3 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity.
Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation.
*4 Only for models with CE marking / UKCA marking on their body.
*5 This is a Class A instrument. This product is intended for use in an industrial environment.
This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*6 This product belongs to Group 1 products. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose.


Options

10-channel multi-point scanner card

For DME1701

20-channel multi-point scanner card

For DME1701

10-channel thermocouple multi-point scanner card

For DME1701

10-channel scanner/10-channel relay card

For DME1701

RS232C card

For DME1701

GPIB card

For DME1701

Kelvin probe (for 4-wire resistance measurement)

4-wire test lead

Catalog

Digital Multimeter – DME1700 Series

https://global.kikusui.co.jp/download/cat_dme1700/

The DME1700 Series digital multimeters have 6½‑digit resolution. This series provides various measurement functions, including voltage, current, resistance, frequency, and temperature, making it suitable for many evaluation and measurement…

Manual

Icon
DME1700

2D CAD

Icon
DME1701
Icon
DME1700

Contact