DC measurement characteristics

ItemDME1700DME1701
DC voltage measurement
(DCV)
Over-rangeFor 100 mV to 100 V ranges20 % over-range allowed
For 1000 V range1 % over-range allowed
A/D converter linearityFor 10 V range0.0001 % of measurement range
Input impedance selection *1For 100 mV to 10 V ranges10 MΩ ± 1 % or >10 GΩ selectable
For 100 V to 1000 V range10 MΩ ± 1 % only
Input bias currentAt 23 °C (73.4 °F)Less than 50 pA
Common mode voltage (Between LO terminal and chassis)Max. 500 Vpk
Auto zero adjustmentWithin ± 1 °C (± 1.8 °F) and within 10 min of the last auto zero adjustmentAdding ± (0.0002 % of measurement range + 3 μV) to accuracy
additional error for disabled settingWithin ± 5 °C (± 9 °F) and within 60 min after the last auto zero adjustmentAdding ± (0.001 % of measurement range + 10 μV) to accuracy
DC current measurement
(DCI)
Over-rangeFor 10 mA to 1 A ranges20 % over-range allowed
For 3 A rangeNo over-range
Common mode current50 Hz or 60 Hz< 600 nApk-pk

*1 Specifications for parallel connection of less than 400 pF.