Voltage/thermometer unit
OP02-PFX
OP02-PFX
110,000円 (税込:121,000円)
- Voltage/thermometer unit
- PFX2512 / 2532
Overview
Features
Voltage/thermometer unit
When monitoring the status of each cell of the battery pack is required, install the optional voltage/ thermometer unit OP02-PFX. By installing OP02- PFX on PFX2512/2532, voltages/temperatures for four cells are able to be monitored/logged with one sheet, respectively. (Up to 3 boards can be installed.) For a battery pack connected in series, monitoring of balance among cells is important. With OP02-PFX, the charge and discharge control can be stopped according to the status of each cell.In addition, it is equipped with a function to stop charge and discharge when the balance beteen the cells in the battery pack becomes large (maximum voltage – minimum voltage). Furthermore, at the time of pulse discharge, voltage can be measured at the same time as the synchronization of all cells for load fluctuations.
Expanded features
- Monitor data: Cell voltage, cell temperature.
- Charge stop conditions: Cell voltage, cell temperature and potential difference among cells
- Discharge stop conditions: Cell voltage and potential difference among cells, cell temperature
- Charge/discharge conditions: Cell voltage, cell temperature, Cell unbalance
- Protective functions: Cell voltage, cell temperature and potential difference among cells
Restricted functions
- The maximum number of channels that 1 unit of personal computer can control is 5 ch.
Specifications
Cell measurement function
Static
OP02-PFX | |
---|---|
Cell voltage | Average voltage of the every 100 ms |
Cell temperature | Temperature measurement function to make thermocouple as temperature detecting element, updated every second |
Cell voltage measurement
Static
OP02-PFX | |
---|---|
Number of measured terminals | 4 |
Measurable range *1 | -2.0000 V to 20.0000 V |
Accuracy *2 | ±(0.05 % of reading + 0.02 % of rating) |
Resolution | 0.1mV |
Measured value | Average voltage of the every 100 ms |
Measurement interval | 100ms |
Cell temperature measurement
Cell temperature measurement *3
OP02-PFX | |
---|---|
Number of measured terminals | 4 |
Thermocouple type | K type |
Measurable range *4 | -100.0 °C to 400.0 °C |
Accuracy *2 *5 | ± 1.5 °C (TYP values) |
Reference junction accuracy *2 *6 | ± 0.5 °C (TYP values) |
Resolution | 0.1℃ |
Measurement interval | 1 s |
*1 You can apply a voltage from -20 V to 22 V. *2 Ambient temperature at 18 °C to 28 °C. *3 The temperature scale conforms to JIS C 1602-1995 (ITS-90). (ITS-90 is an international temperature scale.) *4 Depending on your thermocouple’s specifications (thermocouple class, wire diameter and insulation), the usable temperature range will vary. *5 When the voltage that the thermocouple calibrator produces is measured. *6 This shows the internal sensor performance. This indicates the temperature measurement accuracy of the thermocouple connector. Thermometer accuracy = Measurement accuracy + reference junction compensation + thermocouple tolerance