PXB_General Specifications

ItemPXB20K-50PXB20K-250PXB20K-500PXB20K-1000PXB20K-1500
WeightApprox. 41 kgApprox. 39 kgApprox. 38 kgApprox. 37 kgApprox. 37 kg
DimensionsRefer to Outline Drawing
Environmental conditionsOperating environmentIndoor use, overvoltage category II
Operating temperature0 ℃ to +40 ℃0 °C to +50 °C (32 °F to +122 °F)
Operating humidity20 %rh to 85 %rh (non-condensing)
Storage temperature-25 °C to +60 °C (-13 °F to 140 °F)
Storage humidity90 %rh or less (non-condensing)
AltitudeUp to 2000 m
Cooling methodForced air cooling using fan
Accessories AC INPUT terminal cover, connector kit for external control (1 set), chassis connection wire, DC OUTPUT terminal cover, screws for DC OUTPUT terminal (1 set), EXT SYNC connector cover, SENSING terminal cover, SENSING connector (2 pieces), synchronous operation signal cable kit, For Safety (1 copy), China RoHS sheet (1 sheet), start guide (1 copy), heavy object warning seal (1 sheet)
Withstanding voltageBetween input and GND2200 Vac, 1 minute
Between input and output
Between output and GND500 Vdc, 1 minute1000 Vdc, 1 minute1800 Vdc, 1 minute1800 Vdc, 1 minute3000 Vdc, 1 minute
Insulation resistanceBetween input and GND30 MΩ, 500 Vdc
Between input and output30 MΩ, 500 Vdc30 MΩ, 600 Vdc30 MΩ, 1000 Vdc
Isolation voltage± 250 V± 600 V± 1000 V± 1000 V+2000 V/-1000 V
Electromagnetic compatibility *1 *2 Complies with the requirements of the following directive and standards.
EMC Directive 2014/30/EU
EN 61326-1 (Class A*3)
Safety *1 Complies with the requirements of the following directive and standards.
Low Voltage Directive 2014/35/EU*2
EN 61010-1 (Class I*4, Overvoltage category II, Pollution degree 2*5)
UL 61010-1*6 CAN/CSA -C 22.2 NO.61010-1*6
*1. Does not apply to specially ordered or modified products.
*2. Only for models with CE marking / UKCA marking on their body.
*3. This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided
unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*4. This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded.
*5. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive
pollution will occur except for an occasional temporary conductivity caused by condensation.
*6. Only on models that have cTUVus marking on the panel.