Item/Model | PBZ20-20A | PBZ20-20 | PBZ40-10 | PBZ60-6.7 | PBZ80-5 | ||
---|---|---|---|---|---|---|---|
DC current | Setting range*1 | Bipolar mode | 0.000 A to ±21.000 A | 0.000 A to ±21.000 A | 0.000 A to ±10.500 A | 0.000 A to ±7.035 A | 0.000 A to ±5.250 A |
Unipolar mode | |||||||
Fine feature | ±5 % of rating | ||||||
Setting resolution | 0.001 A (0.0001 A for the fine feature) | ||||||
Setting accuracy*2 | ±0.3 % of rating | ||||||
Temperature coefficient | ±100 ppm/°C of rating (TYP) | ||||||
AC current | Setting range*1 | 0.00 App to 42.00 App | 0.00 App to 42.00 App | 0.00 App to 21.00 App | 0.00 App to 14.07 App | 0.00 App to 10.50 App | |
Setting resolution | 0.01 A | ||||||
Accuracy of setting*3 | ±0.5 % of rating | ||||||
AC frequency | Selectable range | 0.01 Hz to 200.00 kHz | 0.01 Hz to 100.00 kHz | ||||
Setting resolution | 0.01 Hz | ||||||
Setting accuracy | ±200 ppm | ||||||
Sweep | Linear and logarithmic | ||||||
Sweep time | 100 μs to 1000 s (resolution of 100 μs) | ||||||
AC waveform | Type | Sine wave, square wave, triangle wave, and 16 user-defined arbitrary waveforms | |||||
Start phase | 0 ° to 359 ° | ||||||
Square wave duty cycle | 0.1 % to 99.9 % (f < 100 Hz), 1 % to 99 % (100 Hz ≤ f < 1 kHz), 10 % to 90 % (1 kHz ≤ f < 10 kHz), and fixed to 50 % (10 kHz < f) | ||||||
Constant current characteristics | Frequency response *4 | DC to 15 kHz (TYP) | DC to 10 kHz (TYP) | DC to 5 kHz (TYP) | DC to 10 kHz (TYP) | ||
Response *5, *6 | 23 μs, 67 μs, 230 μs, 0.67 ms (TYP) | 35 μs, 100 μs, 350 μs, 1 ms (TYP) | 70 μs, 100 μs, 350 μs, 1 ms (TYP) | 35 μs, 100 μs, 350 μs, 1 ms (TYP) | |||
Overshoot | 5 % or less (TYP) | ||||||
Ripple noise (rms) *7 | 3 mA (TYP) | ||||||
Load effect *8 | ±( 0.01 % of setting + 1 mA ) | ||||||
Source effect *9 | ±( 0.01 % of setting + 1 mA ) |
* 1. The peak value of the sum of the DC current and AC current is limited by the DC current’s settable range. * 2. At an ambient temperature between 18 °C and 28 °C. * 3. At an ambient temperature between 18 °C and 28 °C, with a 100 Hz sine wave, 35 μs response, and shorted output. *4. A frequency where the amplitude ratio of the output current to the external signal input voltage is -3 dB (when the reference frequency is 100 Hz, the response is 35 μs, and a rated load is connected). The frequency response changes according to the load impedance. When the load impedance increases, the frequency response decreases. *5. The rise or fall time (at rated load; excluding when output is turned on and off). The rise and fall times change according to the load impedance. *6. Rise time:The time it takes for the output current to rise from 10 % to 90 % of the rating when the output current is changed from 0 A to the rated current. Fall time:The time it takes for the output current to fall from 90 % to 10 % of the rating when the output current is changed from the rated current to 0 A. * 7. The measurement frequency bandwidth is 10 Hz to 1 MHz (when the output voltage is in the range of 10 % to 100 % of the rated output voltage). * 8. The change in the output current in response to a change in the output voltage from 10 % to 100 % of the voltage rating. * 9. The change in the output current in response to a ±10% change in the input voltage in reference to the nominal input voltage (when the output voltage is in the range of 10 % to 100 % of the voltage rating ).