This table is part of the specifications for:DME1600
Accuracy
±(% of reading + % of range)
The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and slow AC filter (3 Hz to 300 kHz bandwidth).
Measured using a sine wave input whose amplitude is greater than 5% of range.
For the input range from 1 % to 5 %, add 0.1 % of the range (when it is less than 50kHz) or adding 0.13 % of the range (when it is from 50 kHz to 100 kHz)
AC voltage (true rms value)
Range
Resolution
Frequency
1 year (23 °C ± 5 °C)
100.0000 mV
0.1 μV
3 Hz to 5 Hz
1.00+0.04
5 Hz to 10 Hz
0.35+0.04
10 Hz to 20 kHz
0.06+0.04
20 kHz to 50 kHz
0.12+0.05
50 kHz to 100 kHz
0.60+0.08
100 kHz to 300 kHz
4.00+0.50
1.000000 V to 750.000 V
1.0 μV to 1 mV
3 Hz to 5 Hz
1.00+0.03
5 Hz to 10 Hz
0.35+0.03
10 Hz to 20 kHz
0.06+0.03
20 kHz to 50 kHz
0.12+0.05
50 kHz to 100 kHz
0.60+0.08
100 kHz to 300 kHz
4.00+0.50
AC current (true rms value)
Range
Resolution
Frequency
1 year (23 °C ± 5 °C)
1.000000 A
1 μA
3 Hz to 5 Hz
1.00+0.04
5 Hz to 10 Hz
0.30+0.04
10 Hz to 5 kHz
0.10+0.04
3.000000 A
10 μA
3 Hz to 5 Hz
1.10+0.06
5 Hz to 10 Hz
0.35+0.06
10 Hz to 5 kHz
0.15+0.06
Measurement characteristics
Item
Specifications
Overrange
Capable 20 % of the over rage excluding “750 V RMS range”