PLZ-5WH/5WH2 – Firmware Updater

Firmware Updater

This software is to update the firmware version of the PLZ-5WH/5WH2 Series.

Ver.

2.22

Last update

2021/11/11

Configuration

Applied Products: PLZ-5WH/5WH2 Series (Excluding PLZ-5W/5WZ)

[Note]
If your firmware of PLZ-5WH is ver. 2.00, you cannot update to ver. 2.20.
Firstly update to ver. 2.11. Then you can update to ver. 2.20.

Notes

How to use

  • You can update the PLZ5WH/5WH2’s firmware by using a USB memory device.
  • Save the update files (Update.img, CHECKSUM.md5) in the root directory of the USB memory device. Do not change the names of the update files.
  1. Press SYSTEM, Admin, and then Firmware Update.
  2. Insert the USB memory device on which the update files have been saved into the USB port on the front panel, and then press Execute. (An update screen appears.)
  3. Press ENTER.
    Do not turn off the POWER switch while updating is in progress. Do not remove the USB memory device.
  4. After the update complete screen is displayed, turn the POWER switch off and then back on.
    Updating is complete.

The user’s manual for the PLZ-5WH Series (firmware versions 2.1X as of this writing) is still valid after updating to ver. 2.20.

Download

Update to ver. 2.11 only if your firmware of PLZ-5WH is ver. 2.00.

Applied Products: PLZ-5WH Series (Excluding PLZ-5W/5WH2/5WZ)

Release Notes

VER 2.22
2021-10-28

FPGA2.01.0001

  • New Feature Fixed a problem of failing to start at certain ambient temperatures. (IFC2.21.0005: No Change)
VER 2.21
2020-10-23

IFC2.21.0005

  • New Feature Option81 has been added to the Request packet of DHCP.

FPGA2.00.0005

  • Resolved issues The problem of characters "nan" appearing in the measurement has been fixed.
  • Resolved issues The problem where the integrated value was reset when executing the sequence even under a not resetting condition has been fixed.
VER 2.20
2020-02-07

IFC2.20.0000

  • New Feature Added an alarm that goes off when an invalid slave unit is connected.

FPGA2.00.0002

  • Resolved issues Fixed the issue of malfunction caused by the parallel connection of units with imperfect grounding.
VER 2.11
2019-09-12

IFC2.11.0000

  • New Feature Improved to support USB memory devices used for saving or exporting sequence programs.
  • New Feature Added a new function to set a load state after sequence completion.
  • New Feature Added "Cutoff" to the post-trigger source of the sequence function.
  • New Feature Added a reset key (*RST) in "SYSTEM > SCPI" to allow immediate reset.
  • Updated Moved the menu to set the synchronization of sequence start from "SOURCE > More" to "SEQUENCE > More".
  • Updated Changed the factory default setting for soft start time from OFF to 500 μs.
  • Updated Changed the factory default setting for response speed in ARB mode from OFF to 500 μs.
  • Updated Changed the signal pulse width of output from the TRIG OUT connector in pulse operation or sine operation from 1 μs to 10 μs.

FPGA2.00.0001

  • New Feature Improved to support the PLZ-5WH2 series to be connected as a slave unit when the PLZ-5WH series is used as a master unit.