Circuit common (chassis potential) shared by input and output.
3
IN
PM0
Select setup memories and auto test program memories.
4
IN
PM1
5
IN
PM2
6
IN
PM3
7
IN
PM4
8
IN
PM5
9
IN
PM6
10
IN
PM7
11
IN
STB
Recall setup memories and programs selected with the PM0 to PM7 signals.
12
–
Reserved
Not used.
13
–
Reserved
14
–
Reserved
15
IN
START
Start a test.
16
IN
STOP
Stop a test.
17
IN
ENABLE
Enable the START signal.
18
–
COM
I/O circuit common (chassis potential).
19
IN
INTERLOCK-
Activate/release interlock.
20
–
COM
I/O circuit common (chassis potential).
21
–
+24V
+24 V internal power supply output terminal. Maximum output current 100 mA.
22
OUT
H.V ON/LINE ON
Set to on in any of the following conditions. Testing. Auto testing. Voltage remaining across the output terminals. Power being supplied to the EUT from the TOS9303LC through AC LINE OUT.
23
OUT
RISE
Set to on when the voltage is rising.
24
OUT
TEST
Set to on during test time.
25
OUT
PASS
Set to on for the duration of time specified by Pass Hold when a PASS judgment is made.
26
OUT
U FAIL
Set to on continuously when a U-FAIL judgment is made. Or set to on continuously along with the L FAIL signal when CONTACT FAIL judgment is made when a scanner is connected.
27
OUT
L FAIL
Set to on continuously when an L-FAIL judgment is made. Or set to on continuously along with the U FAIL signal when CONTACT FAIL judgment is made when a scanner is connected.
28
–
Reserved
Not used.
29
OUT
READY
Set to on when the product is ready to start a test.
30
OUT
PROTECTION
Set to on when a protection function is activated.
31
OUT
STEP END
Set to on when each step ends during an auto test.
32
OUT
CYCLE END
Set to on when the last step ends during an auto test.
33
OUT
ACW
Set to on when the test mode is set to AC withstanding voltage test.
34
OUT
DCW
Set to on when the test mode is set to DC withstanding voltage test.
35
OUT
IR
Set to on when the test mode is set to insulation resistance test.
36
OUT
EC
Set to on when the test mode is set to earth continuity test.
37
OUT
LC
Set to on when the test mode is set to touch current test or protective conductor test.