TOS5400_Auto Test

ItemSpecifications
Number of programs61
Number of steps2
Test modeThe same settings as for single test settings are possible, except for the following. The same test can be selected consecutively.
  • Limit Voltage is fixed to the maximum rated value per test mode.
  • Test Time is fixed to ON.
  • In IR test, Lower is fixed to ON (only for models that support IR testing).
Fail judgment operationThe following operations are selectable:
  • Continue: All steps are executed even for a FAIL judgment.
  • Abort: Automatic test ends for a FAIL judgment.
Interval time between steps0.1s to 100s
Step start conditionThe following operations are selectable:
  • Immediate: When the INIT or INIT:TEST command is received via remote control, starts all steps consecutively.
  • BUS: Step 1 can be started using the software trigger from remote control. After it is completed, the operation pauses. When the command is then received again, Step 2 starts.
  • External: Step 1 starts upon performing the start operation. After a step is completed, the operation is paused. When you perform another start operation, step 2 starts.
  • Once: When you perform a start operation, all steps are executed consecutively.