TOS5400_Auto Test

This table is part of the specifications for:TOS5401

ItemSpecifications
Number of programs61
Number of steps2
Test modeThe same settings as for single test settings are possible, except for the following. The same test can be selected consecutively.
  • Limit Voltage is fixed to the maximum rated value per test mode.
  • Test Time is fixed to ON.
  • In IR test, Lower is fixed to ON (only for models that support IR testing).
Fail judgment operationThe following operations are selectable:
  • Continue: All steps are executed even for a FAIL judgment.
  • Abort: Automatic test ends for a FAIL judgment.
Interval time between steps0.1s to 100s
Step start conditionThe following operations are selectable:
  • Immediate: When the INIT or INIT:TEST command is received via remote control, starts all steps consecutively.
  • BUS: Step 1 can be started using the software trigger from remote control. After it is completed, the operation pauses. When the command is then received again, Step 2 starts.
  • External: Step 1 starts upon performing the start operation. After a step is completed, the operation is paused. When you perform another start operation, step 2 starts.
  • Once: When you perform a start operation, all steps are executed consecutively.