What precautions should I take when the output terminal of PWR-01 series is not grounded (floating)?

The J1 connector is almost at the same potential as the – (negative) output terminal of PWR-01 series.
Cables and devices connected to this terminal should be rated for an isolation voltage that is equal to or higher than the specified isolation voltage of PWR-01 series.
Even when the J1 connector is not used, be careful not to exceed the isolation voltage.