How do I synchronize the bleeder speed of two PWR-01 series during the master-slave parallel operation?

Set the bleeder circuit setting (CF01) of all connected PWR-01 to the same setting during the master-slave parallel or series operation. The master’s bleeder setting is not reflected in the slave units.