SD007-PFX (BPChecker3000 V1/V2) – Update

Update Module for BPChecker3000 (V1/V2)

This file is the latest updater of “BPChecker3000” for PFX2512/PFX2532. If you currently use the Ver 1, and to use the OP03-PFX, you will need to apply this update. Also, if you want to install this module , the product version “BPChecker3000” must have already been installed, in addition, you will need to update the firmware of PFX2512/PFX2532. Please refer to BPC3000_V2_UpdateGuide_E.pdf for details .

Caution: Please read before updating.

The following notes apply if you are updating Ver 1 to Ver 2.

  1. When you’re using Ver1.0.0.0-Ver1.4.9.0, this update is applied. For a custom-made product, Ver1.5.0.0.-Ver1.9.9.9 can’t use this update.
  2. As a result of functional additions in Ver 2, the test condition file was changed from the data format used in Ver 1. When a test condition file created with Ver 1 is opened with Ver 2, the data is automatically updated for Ver 2. As such, new test conditions created with Ver 2 and test conditions created with Ver 1 that have been opened and saved with Ver 2 cannot be used with Ver 1. We recommend that you back up test conditions created with Ver 1 in a separate folder or the like.
  3. Ver 2 is designed so that test conditions acquired with Ver 1 can be viewed with Graphviewer of Ver 2, but problems may occur depending on the content of test data, the PC performance, and so on. If a problem occurs such as not being able to open test conditions of Ver 1 after updating to Ver 2, it is possible to revert temporarily to Ver 1. See BPC3000_V2_UpdateGuide_E.pdf.
  4. Ver 2 does not support Windows XP.
  5. The software environmental settings (graph axis, color, etc.) are passed on from Ver 1, but some settings may be reset to default values.

Ver.

2.0.7.00

Last update

2023/3/9

Configuration

Notes

When using GPIB-connected temperature chambers on Win 7/8 x64 environment, 64bit versions of GPIB driver and VISA library are required.

Download

Release Note

VER 2.0.7.00
2023-2-14

PFX firmware

Resolved issues
Fixed a problem that may not work with CV charging when charging CC-CV using a battery with high impedance.
Resolved issues
Fixed a problem that only the maximum cell voltage is not displayed in the test result when OP02-PFX is installed.
Resolved issues
Fixed a problem that an incorrect mode is displayed (recorded) when operating in cell CV mode.
VER 2.0.6.02
2021-1-20

PFX firmware / BPC3000

Resolved issues
Fixed a problem that activated the overvoltage protection (OVP) upon completion of the charge in combination with PWR-01 series.
Resolved issues
Fixed a problem when the test was paused and resumed in the last profile, only one profile was excecuted to complete the test regardless of the resuming position.
Resolved issues
Fixed a problem that decreased the detection accuracy of the overcharge capacity protection (OAH) (the OAH detection value (capacity: Ah) was rounded down to the nearest integer).
Improvement
Fixed a problem that required the time to switch profiles under the test condition with many profiles depending on the PC environment.
VER 2.0.5.00
2020-9-03

PFX firmware

Resolved issues
Fixed a problem that caused an error due to false detection of the option board when the option board was not inserted.
VER 2.0.4.00
2018-6-01

BPC3000

New Feature
Added support for model Ids that combine power supplies (PWR-01) and electronic loads (PLZ-5W).
New Feature
Added the “Save the data for Graph Viewer” option to TestExecutive. When the check box is cleared, memory consumption is suppressed, which makes long-term profile operation possible.
Improvement
Improved the memory usage algorithm and processing speed when a profile is run over a long time period on TestExecutive.
Resolved issues
Fixed a problem that was not applying selections correctly when a profile with the Delta Time of Data Record Method set to Fix Time is edited on Test Condition Editor.
VER 2.0.3.00
2017-12-01

PFX firmware / BPC3000

Resolved issues
Fixed a problem that was displaying an incorrect error code for electronic load alarms.
Resolved issues
Fixed a problem that the profile template created before V 2.01 is not displayed.
Resolved issues
Fixed a problem that was resetting the unbalanced margin display on the profile screen.
VER 2.0.2.12
2017-1-23

PFX firmware / BPC3000

Resolved issues
Fixed a problem that was always causing the test to stop with an OTP when I-V characteristic Prof was executed immediately after starting the PFX. (This problem could be reproduced only when a thermometer was not connected to the PFX.)
New Feature
Added support for the ID7306 model.
Improvement
Improved the PFX search algorithm on TestExecutive and IoConfig.
Resolved issues
Refined the model ID file loading process.
Resolved issues
Fixed a problem that was always causing the cell CV to operate during CC-CV charge Prof or pattern Prof and preventing current from flowing when all cell voltage measurements were turned off when an OP board was installed.
Resolved issues
Fixed a problem that causing an exception when no channels could be connected during TestExecutive startup.
VER 2.0.1.03
2016-7-29

PFX firmware / BPC3000

Resolved issues
Fixed a problem that was causing the cell CV function to operate when a given condition was met even when there was no OP board.
Resolved issues
Fixed a problem that was preventing programs from proceeding to the next loop if “RestTemp (LastStep)” is enabled in a pattern profile or I-V characteristic profile when there was no thermometer connected (e.g., OPEN).
Improvement
Redesigned the new text output format. (The two-row column header was changed to one.)
Improvement
Redesigned the GraphViewer text export function so that the elapsed time column is changed to the total time column.
New Feature
Added a preset timeout function for temperature chamber synchronization.
Improvement
Changed the supported KI-VISA version to v5.5.0. (To use temperature chambers, update the KI-VISA version to v5.5.0 or later.)
VER 2.0.0.38
2016-1-06

Global / BPC3000

New Feature
Added a cell CV function.
New Feature
Voltage measurement of up to 72 cells is possible by connecting SL01-PFXs.
Improvement
Drastically improved the operability of Graph Viewer. [Description] • Added an auto creation wizard for I-V characteristic graphs. • Printing can be executed from the shortcut menu. • Improved the operability of moving and zooming data lines. • Redesigned the pane configuration. • Redesigned the shortcut menu. • Redesigned button arrangement.
Improvement
Drastically improved the print content of Graph Viewer. [Description] • Improved the smudging of graphs during printing. • Redesigned grayscale printing. • Redesigned the print format of test condition and the like.

Ver 1.4.9.00

VER 1.4.9.00
2015-5-27

PFX2512 / BPC3000

Resolved issues
Fixed a problem that was causing the start of the next test to be held until the stop time expires even when the test itself has stopped earlier.
Resolved issues
Expanded the measurable temperature of the PFX thermometer from down to -40°C to down to -50°C.
Improvement
Updated the model ID profile to suppress the disparity between electronic load devices.
VER 1.4.8.01
2014-10-06

BPC3000

Resolved issues
Fixed a problem with the IT functionality not operating correctly.
Resolved issues
Fixed a problem that was preventing the 0.1 mV and 0.1 mA accuracies from being achieved in high-speed sampling mode.
Improvement
Fixed so that VISA resources LAN and GPIB can be selected in addition to serial.
Resolved issues
Fixed other problems.
VER 1.4.7.97
2013-7-09

BPC3000

Resolved issues
Changed the description of the chamber connection method in the operation manual.
VER 1.4.7.94
2013-6-06

BPC3000

Improvement
Improved the update feature so that users using special-order products can also continue to use their environment
VER 1.4.6.83
2013-3-10

PFX2512 / Test Executive / Test Condition Editor / GraphViewer

Resolved issues
Fixed the overshoot that was occurring the first time the PAT output was turned on.
Resolved issues
Fixed an error in the default current control method of pattern charge/discharge.
Improvement
Added a default time-based monitoring feature for pattern charge/discharge.
Resolved issues
Fixed a problem in the handling of the limit current in CP charge/discharge.
Improvement
Added support for 2-port LAN.
Improvement
Added a feature that saves the manually input IP address.
Resolved issues
Fixed an error in the CV value for CC-CV charge and CC-CP discharge in the profile master.
Improvement
Optimized the profile master.
Resolved issues
Fixed a problem that was causing incorrect processing when the edit button was processed if the profile master was empty.
Resolved issues
!!!DELETE!!!
Resolved issues
Fixed a problem that was causing the extension of life text file names to be changed.
VER 1.4.3.57
2013-1-31

Test Executive

Resolved issues
Fixed a problem that was preventing current graphs greater than or equal to 50 A from being displayed.
VER 1.4.2.56
2013-1-08

BPC3000 / Test Condition Editor / IO Config / Test Executive

Improvement
Added model ID 7119.
Resolved issues
Fixed a problem that was causing the application to close when model ID 7114 was selected.
Improvement
Prohibited the voltage range to be switched from IoConfig.
Resolved issues
Fixed a problem in which the extension of life result file names was lacking the dot.
Improvement
Changed the default chamber humidity setting to -1.
Resolved issues
Fixed a problem that was causing the low voltage range to return to high when a re-test was performed.
Resolved issues
Fixed an error in the updating of the test condition grid of TestExecutive.
VER 1.4.1.36
2012-11-02

BPC3000

Resolved issues
Fixed a problem that occurred when the chamber humidity was not set.
Resolved issues
Added model ID 7114.
VER 1.4.0.35
2012-10-22

BPC3000 / PFX2512

Improvement
Added an English help document.
Resolved issues
Fixed a problem in the trigger specification of pattern profiles.
VER 1.3.6.33
2012-8-22

GraphViewer / Test Condition Editor

Resolved issues
Fixed a problem that failed to show two profiles in the maximum and minimum displays of the cell voltage.
Resolved issues
Fixed a problem that was causing the time to drift when the cell voltage was expanded time-wise.
Resolved issues
Fixed a problem with the VI cursor movement.
Resolved issues
Fixed a problem with the total number of steps in patterns.
VER 1.3.3.28
2012-7-20

GraphViewer / PFX2512

Resolved issues
Fixed a problem with the processing of the Next button when option board data was displayed.
Resolved issues
Fixed a problem that was causing an exception when a BPCG3 file was opened directly and the content was zoomed.
Resolved issues
Fixed a problem that was showing abnormal values in the result data when the cell was excluded from measurements.
Resolved issues
Fixed a problem that was causing pattern charge/discharge execution times to be reduced in 60 Hz regions.
VER 1.3.2.26
2012-6-21

Test Condition Editor / GraphViewer

Resolved issues
Fixed a problem with the power limit when test condition files created on a special-order power supply model was being re-edited.
Resolved issues
Fixed a problem with the processing of the Next button when option board data was displayed.
VER 1.3.0.20
2012-4-24

Test Executive / PFX2512

Resolved issues
Fixed a problem that was causing an alarm when tests were executed without configuring the option board.
Improvement
Ver1.0.2 Changed the humidity calibration point.
VER 1.1.0.12
2012-3-27

Test Condition Editor / GraphViewer

Improvement
Added a profile master check and repair feature.
Improvement
Changed so that the X-axis can be set for each graph.

Please feel free to contact us with any questions or concerns.

Download a Product CatalogView List
Request a QuotationContact Us