(確認用)KHA3000

Published: August 2022 Please check the publication date of this article. The information on products, pricing, and technology may not reflect the latest updates.

Ver 3.61
2026-9-03

KHA3000 firmware

Improvement
Added support for the harmonic current limit standard IEC 61000-3-2 Ed5.2 with the following changes. brbr- Discontinued support for IEC 61000-3-2 Ed5.0. Accordingly, the standard selection option has been changed from IEC 61000-3-2 Ed5.0 to IEC 61000-3-2 Ed5.2. The displayed local standard name has also been changed from EN IEC 61000-3-2(2019) to EN IEC 61000-3-2(2019)/A2(2024). brbr- Made the following additional changes in line with the update from IEC 61000-3-2 Ed5.0 to Ed5.2: brbr - Updated the POHC calculation method to the method specified in Annex C.3. brbr - Changed the Class C limit value for the third harmonic from “30*λ[%]” to “27[%].” brbr - Applied positive and negative peak voltage judgment (crest factor judgment) to Classes A and B in “Checking the AC power supply.” brbr - Updated the threshold values used for repeatability check as follows: brbr 11th order or lower: The average harmonic current being compared must be within ±(5% of the limit value + 1 mA). brbr 12th order or higher: The average harmonic current being compared must be within ±(10% of the limit value + 1 mA). brbr - Changed the current values and POHC values printed on reports to the average values during the test period. (In earlier versions, the maximum values during the test period were printed.) brbr - Changed the exclusion condition for harmonic current judgment to less than the greater of 0.6 % of the average input current or 5 mA. (In earlier versions, the condition was less than the greater of 0.6 % of the maximum input current or 5 mA.) brbr - Added “IEC 61000-3-2 Ed5.2” as a parameter for the command “HARM:STAN:LIM”. brbr - Added “IEC 61000-3-2 Ed5.2” to the response for the command “HARM:STAN:LIM:CAT?”. brbr - Added “EN IEC 61000-3-2(2019)/A2(2024)” to the default response for the command “HARM:STAN:LIM:ALI:NAME?”.
Improvement
Added support for the harmonic current limit standard IEC 61000-3-12 2021 with the following changes. brbr- Discontinued support for IEC 61000-3-12 2011. Accordingly, the standard selection option has been changed from IEC 61000-3-12 2011 to IEC 61000-3-12 2021. The displayed local standard name has also been changed from EN 61000-3-12(2011) to EN 61000-3-12(2011)/A1(2024). brbr- Added a notification that is displayed after the test when the following settings and conditions are met: classification of EUT is “Balanced 3-P,” and the limit value is either “Spec Bal’d 3-P” or “Table5 3-P.” brbr- Added report output for conformance against the condition (c) or (f) when the following settings are met: classification of EUT is “Balanced 3-P,” and the limit value is either “Spec Bal’d 3-P” or “Table5 3-P.” brbr- Added “IEC 61000-3-12 2021” as a parameter for the command “HARM:STAN:LIM”. brbr- Added “IEC 61000-3-12 2021” to the response for the command “HARM:STAN:LIM:CAT?”. brbr- Added “EN 61000-3-12(2011)/A1(2024)” to the default response for the command “HARM:STAN:LIM:ALI:NAME?”. brbr- Added the commands “HARM:IEC312:COND:RES?” and “HARM:IEC312:COND:COUN?”.
Improvement
Added support for the flicker limit standard IEC 61000-3-3 Ed3.2 with the following changes. brbr- Discontinued support for IEC 61000-3-3 Ed3.1. brbr- Changed the standard name printed in reports from IEC 61000-3-3 Ed3.1 to IEC 61000-3-3 Ed3.2. The displayed local standard name has also been changed from EN 61000-3-3(2013)/A1(2019) to EN 61000-3-3(2013)/A2(2021). brbr- Added “EN 61000-3-3(2013)/A2(2021)” to the default response for the command “VFL:STAN:LIM:ALI:NAME?”.
Improvement
Changed the test results file format; test results files from versions 3.55 and earlier can no longer be loaded.
Improvement
Enabled screen hard copies to be taken while a dialog box is displayed.
Resolved issues
Corrected the standard name printed in reports for IEC 61000-3-3 (Pst Auto/Manual) so that it changes according to the Measuring Technique Standard setting. brbr- When IEC 61000-4-15 Ed1.1 is selected brbr"EN61000-3-3(2008) brbr*IEC61000-3-3:Ed2.0(2008)" brbr- When IEC 61000-4-15 Ed2.0 is selected brbr"EN61000-3-3(2013)/ A2(2021) brbr*IEC61000-3-3:Ed3.2(2021)"
Resolved issues
Corrected the standard name printed in reports for IEC 61000-3-11 Ed2.0 as follows. brbrBefore: “IEC61000-3-11(2019)” brbrAfter: “EN IEC61000-3-11(2019)”
Resolved issues
Resolved an issue that prevented the following content from displaying correctly when loading a test results file for IEC 61000-3-12 2021 (IEC 61000-3-12 2011 in versions 3.55 and earlier). brbr- 2D Harmonics brbr- 3D Harmonics brbr- Harmonics List
Resolved issues
Resolved an issue that caused an error when the ASSIST key was pressed on the “Checking the AC power supply” display for IEC 61000-3-12 2004 and IEC 61000-3-12 2021 (IEC 61000-3-12 2011 in versions 3.55 and earlier).