SD009-PCR-LE/WE

Power Line Disturbance Immunity Testing Software

immunity

80,000 (税込:88,000

  • Capable of performing compliance testing applied to the IEC61000-4-11, 4-13, 4-14, 4-17, 4-27, 4-28, 4-29, 4-34 or to perform preliminary tests before the standard testing.
  • Applied to single-phase/ single-phase three-wire/ three-phase and the DC input power.
  • Not only for compliance standard testing, it can be also employed for advance checking in development phases and for margin tests.
  • Sequence chain (up to 10 sequences) features improve test efficiency
  • Waveform preview provides a general idea of the output wave-form without having an oscilloscope
  • The setting condition file and the test result file can be saved in txt or csv format.

What is myKIKUSUI?

カート機能が動くまでの
SKU: immunity
  • List of conformance to the EMC standard tests
  • Operating environment
  • Downloads

List of conformance to the EMC standard tests

The latest standards for IEC61000-4 supported!

SD009-PCR-LE/WE (Product name: Quick Immunity Sequencer 2) is an application software to perform immunity tests based on the standard for power line disturbance in the immunity test standard of EMC (IEC61000-4 series). Not only can it be used for compliance testing based on the latest standards or for some types of preliminary testing, but the software can be also employed for advance checking in development phases and for immunity margin tests, because it allows extended testing conditions to be set as needed.

✓:Conforming as standard △:Partially non-conforming -:Function not available

*Immunity testing for units with 16 A/phase except for those required by IEC61000-4-34

  • StandardItemConforming
    Single-phaseThree-phase
    IEC61000-4-11
    Voltage dipping, instantaneous power failure and voltage variation
    Voltage drop (dip)*1*1
    Instantaneous power failure*1*1
    Voltage variation
    IEC61000-4-13
    Higher harmonics wave/interharmonic wave
    Flat curve
    Over swing
    Frequency sweep
    Odd harmonics the order of which is not a multiple of 3
    Odd harmonics the order of which is a multiple of 3
    Even harmonics
    Interharmonics
    Meister curve
    IEC61000-4-14
    Voltage swing
    Voltage swing
    Interval
    IEC61000-4-17
    Ripple at the DC input power terminal
    Single-phase rectifier circuit
    Three-phase rectifier circuit
    IEC61000-4-27
    Unbalance in units
    Unbalance*2
    IEC61000-4-28
    Variation in-power supply frequency for units with 16 A/phase
    Frequency variation
    IEC61000-4-29
    Voltage drop (dip), instantaneous power failure
    and voltage variation in DC
    Voltage drop (dip)
    Instantaneous power failure*3
    Voltage variation
    IEC61000-4-34
    Voltage drop (dip), instantaneous power failure and voltage variation for units with input current exceeding 16 A/phase
    Voltage drop (dip)*4*4
    Instantaneous power failure*4*4
    Voltage variation
    • *1 Conforms to the standard when used in combination with IEC Dip Simulator DSI series. If using the PCR-WEA/WEA2 alone, the voltage dips and short-time power failures are preliminary tests.
    • *2 110%, 95.2%, 93.5%, 90%, 87%, 80%, 74%, 71%, 66% need to respond to sudden changes of 1 μs to 5 μs. The voltage response of PCR-WEA/WEA2 is more than 40 μs at FAST, which is a preliminary test.
    • *3 Must support output impedance greater than 100 kΩ. The PCR-WEA/WEA2 output impedance is less than 100 kΩ and therefore designed for preliminary testing purposes. Only PCR12000WEA2R conforms to the standards.
    • *4 The device between the range of 16A to 75 A requires having the capability of rapid change with 1 μs to 5 μs. The device exceeding 75 A is not required to have the capability of rapid change with 1μs to 5 μs.(It is relaxed to 1 μs to 50 μs for the device exceeding 75 A.)

  • StandardItemConforming
    Single-phaseThree-phase
    IEC61000-4-11
    Voltage dipping, instantaneous power failure and voltage variation
    Voltage drop (dip)*1*1
    Instantaneous power failure*1*1
    Voltage variation
    IEC61000-4-13
    Higher harmonics wave/interharmonic wave
    Flat curve
    Over swing
    Frequency sweep
    Odd harmonics the order of which is not a multiple of 3
    Odd harmonics the order of which is a multiple of 3
    Even harmonics
    Interharmonics
    Meister curve
    IEC61000-4-14
    Voltage swing
    Voltage swing
    Interval
    IEC61000-4-17
    Ripple at the DC input power terminal
    Single-phase rectifier circuit
    Three-phase rectifier circuit
    IEC61000-4-27
    Unbalance in units
    Unbalance*2
    IEC61000-4-28
    Variation in-power supply frequency for units with 16 A/phase
    Frequency variation
    IEC61000-4-29
    Voltage drop (dip), instantaneous power failure
    and voltage variation in DC
    Voltage drop (dip)
    Instantaneous power failure*3
    Voltage variation
    IEC61000-4-34
    Voltage drop (dip), instantaneous power failure and voltage variation for units with input current exceeding 16 A/phase
    Voltage drop (dip)*4*4
    Instantaneous power failure*4*4
    Voltage variation
    • *1 Conforms to the standard when used in combination with DIP Simulator. If using the PCR-LE/LE2 alone, the voltage dips and short interruptions are preliminary tests.
    • *2 Capability of rapid change with 1 μs to 5 μs is required for 110 %, 95.2 %, 93.5 %, 90 %, 87 %, 80 %, 74 %, 71 %, 66 %. Preliminary test is capable since the voltage response of the PCR-LE/LE2 is 20 μs in FAST mode and 30 μs in MEDIUM mode.
    • *3 Must support output impedance greater than 100 kΩ. The PCR-LE/LE2 output impedance is less than 100 kΩ and therefore designed for preliminary testing purposes.
    • *4 The device between the range of 16 A to 75 A requires to have the capability of rapid change with 1 μs to 5 μs. The device exceeding 75 A does not require to have the capability of rapid change with 1 μs to 5 μs. (It is relaxed to 1 μs to 50 μs for the device exceeding 75 A.)

  • StandardItemConforming
    Single-phaseThree-phase
    IEC61000-4-11
    Voltage dipping, instantaneous power failure and voltage variation
    Voltage drop (dip)*1*1
    Instantaneous power failure*1*1
    Voltage variation
    IEC61000-4-13
    Higher harmonics wave/interharmonic wave
    Flat curve
    Over swing
    Frequency sweep
    Odd harmonics the order of which is not a multiple of 3
    Odd harmonics the order of which is a multiple of 3
    Even harmonics
    Interharmonics
    Meister curve
    IEC61000-4-14
    Voltage swing
    Voltage swing
    Interval
    IEC61000-4-17
    Ripple at the DC input power terminal
    Single-phase rectifier circuit
    Three-phase rectifier circuit
    IEC61000-4-27
    Unbalance in units
    Unbalance*2
    IEC61000-4-28
    Variation in-power supply frequency for units with 16 A/phase
    Frequency variation
    IEC61000-4-29
    Voltage drop (dip), instantaneous power failure
    and voltage variation in DC
    Voltage drop (dip)*3
    Instantaneous power failure*4
    Voltage variation
    IEC61000-4-34
    Voltage drop (dip), instantaneous power failure and voltage variation for units with input current exceeding 16 A/phase
    Voltage drop (dip)*5*5
    Instantaneous power failure*5*5
    Voltage variation
    • *1 Conforms to the standard when used in combination with IEC Dip Simulator DSI series. If using the PCR-WE/WE2 alone, the voltage dips and short-time power failures are preliminary tests.
    • *2 110 %, 95.2 %, 93.5 %, 90 %, 87 %, 80 %, 74 %, 71 %, 66 % need to respond to sudden changes of 1 μs to 5 μs. The voltage response of PCR-WE/WE2 is more than 55 μs at FAST, which is a preliminary test.
    • *3 The voltage response of PCR-WE/WE2 is more than 55 μs at FAST, which is a preliminary test.
    • *4 Must support output impedance greater than 100 kΩ. The PCR-WE/WE2 output impedance is less than 100 kΩ and therefore designed for preliminary testing purposes.
    • *5 The device between the range of 16A to 75 A requires having the capability of rapid change with 1 μs to 5 μs. The device exceeding 75 A is not required to have the capability of rapid change with 1μs to 5 μs. (It is relaxed to 1 μs to 50 μs for the device exceeding 75 A.)

Operating environment

  • A PC with Microsoft Windows1 10(x86/x64) or 7(x86/x64)
  • Microsoft.NET Framework 2.0
  • 8 GB or more of RAM (16 GB or more recommended)
  • 1366×768 dots or higher resolution
  • 100 MB or more of free hard disk space (additional space is required to save data)
  • CD-ROM drive
  • Mouse or other pointing device
  • VISA library : NI-VISA 5.0.3 or later, Keysight IO Libraries Suite 16.1 or later, or KI-VISA 5.0.5 or later
  • USB cable (only when using the USB interface)
  • GPIB board and IEEE488 cable (only when using the GPIB)
  • Crossover serial cable (only when using the RS-232C interface)
  • Straight type LAN cable (only when using LAN with HUB connection)
  • Cross-type LAN cable (only when using LAN with one-to-one connection)

Downloads

Operation Manuals

SD009-PCR-LE/WE
Icon

    Updater / Driver

    Please feel free to contact us with any questions or concerns.

    Download a Product Catalog View List
    Request a Quotation Contact Us
    close

    Search