Battery Impedance Meter

BIM1000 Series

BIM1000 series

260,000420,000

  • Maximum voltage measurement: 1000V max. (BIM1100), 300V max. (BIM1030)
  • Voltage measurement accuracy: ±(0.01% of reading +3 digit)
  • Resistance measurement accuracy: ±(0.5% of reading +5 digit)
  • Resistance measurement ranges: 3mΩ / 30mΩ / 300mΩ / 3Ω
  • High resolution: Voltage 10μV (6V range), Resistance 0.1μΩ (3mΩ range)
  • Measurement frequency: 1kHz ±0.2Hz
  • Sampling speed (Resistance & voltage simultaneous measure-ments): 20ms (FAST)
  • Zero Adjustment Function: Effective for decreasing measure-ment error.
  • Measurement logging (500 pairs) and collective transfer function
  • SIGNAL I/O, RS232C and USB as standard interface (BIM1030 WITH LAN and BIM1100 WITH LAN are standard equipped also with a LAN port)
  • New high visibility color display.

LAN LXIUKCA
ImageSKU標準価格WEBかんたんお見積り
BIM1030±300 Vwith LAN260,000
(税込:286,000
BIM1030 with LAN±300 Vwith LAN280,000
(税込:308,000
BIM1100±1000 Vwith LAN400,000
(税込:440,000
BIM1000 series±1000 Vwith LAN420,000
(税込:462,000
SKU: BIM1000 series
  • Features
  • Options
  • Downloads

Outline

An ideal unit for manufacturing and inspection applications of rechargeable batteries.

Ever-changing battery technology requires batteries powering electric vehicles to have high voltage, high power and low impedance. The Battery Impedance Meter, or BIM1000 Series, is capable of measuring up to 1000 V of test voltage for simultaneous measurements of both battery voltage and resistance at high speeds. The BIM is the ideal equipment for power battery development research and production tests.

Applications

Color liquid crystal display (LCD) / Comparator Functions

Color liquid crystal display (LCD)

High visibility color monitor. The resistance, voltage, upper and lower limits values are displayed at a glance.

Comparator Functions

The comparator functions allows setting HIGH/LOW, and resistance and voltage can be simultaneously judged by independent comparators. Judgment results are provided on the display. External I/O is available for signal output.

Downloads

Product Catalog

Icon
Battery Impedance Meter - BIM1000 Series

Operation Manuals

Options

Line-up

ImageSKU標準価格WEBかんたんお見積り
BIM1030±300 Vwith LAN260,000
(税込:286,000
BIM1030 with LAN±300 Vwith LAN280,000
(税込:308,000
BIM1100±1000 Vwith LAN400,000
(税込:440,000
BIM1000 series±1000 Vwith LAN420,000
(税込:462,000

Overview

An ideal unit for manufacturing and inspection applications of rechargeable batteries.

Ever-changing battery technology requires batteries powering electric vehicles to have high voltage, high power and low impedance. The Battery Impedance Meter, or BIM1000 Series, is capable of measuring up to 1000 V of test voltage for simultaneous measurements of both battery voltage and resistance at high speeds. The BIM is the ideal equipment for power battery development research and production tests.

Production line testing, etc.

Color liquid crystal display (LCD) / Comparator Functions

color LCD

High visibility color monitor. The resistance, voltage, upper and lower limits values are displayed at a glance.

Comparator Functions

The comparator functions allows setting HIGH/LOW, and resistance and voltage can be simultaneously judged by independent comparators. Judgment results are provided on the display. External I/O is available for signal output. Applications

Specifications

Voltmeter (The range can be AUTO setting available)
Open in a new windowZOOM
ItemBIM1030 / BIM1100
Rated inputBIM1030 : ± 300V / BIM1100 : ± 1000V
Range6V60V300V(BIM1030)600V(BIM1100)1000V(BIM1100)
Maximum display value *1±6.30000V±63.0000V±315.000V± 630.000V±1050.00V
Resolution10μV100μV1mV10mV
Accuracy *2± ( 0.01% of reading + 3digit )
Temperature coefficient± ( 0.001% of reading + 0.3digit ) / ℃
Response time *3Approx. 1 ms

*1. Displays OVER when the measurement range is exceeded. *2. Add ±2 digit when the sampling speed is set to FAST or MID. *3. The time for the product’s measurement circuit to stabilize when a probe in an open state is connected to the DUT.

Resistance meter (The range can be AUTO setting available)
Open in a new windowZOOM
ItemBIM1030 / BIM1100
Measurement methodFour-terminal measurement method
Range3mΩ30mΩ300mΩ
Maximum display value *13.1000mΩ31.000mΩ310.00mΩ3.1000Ω
Resolution0.1μΩ1μΩ10μΩ100μΩ
Measured current *2100mA10mA1mA
Measurement frequency1kHz ± 0.2Hz
Accuracy *3±( 0.5% of reading + 5digit)
Temperature coefficient± ( 0.05% of reading + 1digit ) / ℃± ( 0.05% of reading + 0.5digit) / ℃
Response time *4Approx. 2 ms

*1. Displays OVER when the measurement range is exceeded. *2. Within error ± 10 %. *3. Add ±3 digit when the sampling speed is set to FAST and ±2 digit when the sampling speed is set to MID. *4. The time for the product’s measurement circuit to stabilize when a probe in an open state is Environ- ment Power supply Spec guaranteed range Operating range Storage range Temperature Humidity Temperature Humidity Temperature Humidity connected to the DUT.

Sampling time
Open in a new windowZOOM
ItemBIM1030 / BIM1100
Sampling speedFASTMEDIUMSLOW
Power supply frequency50Hz20ms50ms160ms
±60Hz42ms150ms
Judgment feature
Open in a new windowZOOM
ItemBIM1030BIM1100
Judgment methodWindow comparator method. Judgment made with software.
ResistanceSetting range0.0001Ω 〜 3.1000Ω
Resolution100μΩ
VoltageSetting range0.000V 〜 315.000V0.00V 〜 1050.00V
Resolution1mV10mV
Interface
Open in a new windowZOOM
ItemBIM1030 / BIM1100
RS232CD-SUB 9-pin connector, EIA-232-D compliant
USBComplies with USB Specification 2.0. Data rate: 12 Mbps max. (Full Speed)
Complies with USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0
LAN *1IEEE 802.3, 100Base-Tx/10Base-T Ethernet, IPv4, RJ45, Auto-negotiation supported but Auto MDI/MDI-X not supported.
SIGNAL I/OD-SUB 25-pin connector.

*1. BIM1030 WITH LAN or BIM1100 WITH LAN only

General specifications
Open in a new windowZOOM
ItemBIM1030BIM1100
EnvironmentInstallation locationIndoors, 2000 m or less
Spec guaranteed rangeTemperature / Humidity18 °C to 28 °C (-4 °F to 158 °F) / 20 %rh to 80 %rh (no condensation)
Operating rangeTemperature / Humidity0 °C to 40 °C (32 °F to +122 °F) / 20 %rh to 80 %rh (no condensation)
Storage rangeTemperature / Humidity-10 °C to 60 °C (-4 °F to 158 °F) / 90 %rh or less (no condensation)
Power supplyInput voltage range / Input frequency range85 Vac to 264 Vac (100 Vac to 240 Vac)/47 Hz to 63 Hz
Rated power30VA
Isolation voltage± 300 V max± 1000 V max
Insulation resistance30 MΩ or more (500 Vdc)(between AC LINE and chassis)
Withstanding voltageBetween the AC LINE and the chassis1500 Vac for 1 minute, 10 mA or less
Between all the mea- surement terminals and the chassis2000 Vdc for 1 minute, 1 mA or less
Between all the mea- surement terminals and SIGNAL I/O2000 Vdc for 1 minute, 1 mA or less
External dimensions / Weight214(8.43)W×80(3.15)H×300(11.81)D mm(inches) (Does not include protrusions)/ Approx. 3 kg (6.6 lbs)
Electromagnetic compatibility (EMC) *1 *2Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3
Safety *1Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2 EN 61010-1 (Class I *5, Pollution Degree 2 *6), EN 61010-2-030, EN 61010-031

*1. Does not apply to specially ordered or modified products. *2. Only on models that have CE/UKCA marking on the panel. *3. This is a Class A instrument. This product is intended for use in an industrial environment. This prod- uct may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4. This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequen- cy energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5. This is a Class I instrument. Be sure to ground this product’s protective conductor terminal.The safety of this product is guaranteed only when the product is properly grounded. *6. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollu- tion will occur except for an occasional temporary conductivity caused by condensation.